FDA recall record

ZOLL Medical OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 recall — FDA Z-0977-2018

A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult…

Class IIclassification
Z-0977-2018official record ID
Mar 21, 2018report date

Recall record at a glance

Official recordZ-0977-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZOLL Medical Corporation
Report dateMar 21, 2018
Recall initiation dateAug 4, 2017
Quantity3400
LocationChelmsford, MA, United States

Official product description

OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 01 (case of 8 single electrodes, UDI: 10847946016279) and 8900 0224 01 (single electrode, UDI: 00847946016272) Product Usage: Intended Use: Defibrillation, Cardioversion, Noninvasive Pacing, EGG Monitoring, CPR Sensor. For use with ZOLL Defibrillators: R Series, M Series. By Trained Personnel only, Including: Physicians, Nurses, Paramedics, Emergency Medical Technicians, Cardiovascular Laboratory Technicians.

Why the product was recalled

A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult energy levels when used with the ZOLL R Series Defibrillator. Defective electrodes will display PEDIATRIC PADS IN USE on the R Series defibrillator. Using the affected product with the M Series Defibrillator poses no risk.

Distribution information

Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and in the countries of Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1321-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1319-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1320-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1322-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1316-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1317-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation