FDA recall record

Life Technologies Oncomine Dx Target Test recall — FDA Z-1480-2018

The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false…

Class IIclassification
Z-1480-2018official record ID
May 2, 2018report date

Recall record at a glance

Official recordZ-1480-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLife Technologies Corporation
Report dateMay 2, 2018
Recall initiation dateFeb 8, 2018
Quantity30
LocationFrederick, MD, United States

Official product description

Oncomine Dx Target Test, (Model No. A32451), Oncomine Dx Target RNA/DNA Panel (Model No. A32441). A qualitative In Vitro Diagnostic test.

Why the product was recalled

The lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment regimen.

Distribution information

Distributed to states: AZ, CA, NC, TX, and VA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records