FDA recall record

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R recall — FDA Z-1264-2022

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function…

Class IIclassification
Z-1264-2022official record ID
Jun 22, 2022report date

Recall record at a glance

Official recordZ-1264-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateJun 22, 2022
Recall initiation dateApr 20, 2022
Quantity4 units
LocationCenter Valley, PA, United States

Official product description

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R

Why the product was recalled

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Distribution information

CA FL MA MD MI NJ OH PR WI

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.