FDA recall record

Olympus Scientific Solutions Americas Corporation (OSSA) VANTA XRF Analyzer recall — FDA Z-1036-2017

It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at…

Class IIclassification
Z-1036-2017official record ID
Feb 8, 2017report date

Recall record at a glance

Official recordZ-1036-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOlympus Scientific Solutions Americas
Report dateFeb 8, 2017
Recall initiation dateJan 7, 2017
QuantityUS 74
LocationWaltham, MA, United States

Official product description

Olympus Scientific Solutions Americas Corporation (OSSA) VANTA XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems

Why the product was recalled

It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.

Distribution information

US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.