FDA recall record

Integrated Medical Systems Olympus, Model No. GIF-Q180,flexible scope, large diameter, video recall — FDA Z-0894-2016

If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become…

Class IIclassification
Z-0894-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0894-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegrated Medical Systems Inc.
Report dateMar 2, 2016
Recall initiation dateJan 20, 2016
Quantity185 instruments
LocationBirmingham, AL, United States

Official product description

Olympus, Model No. GIF-Q180,flexible scope, large diameter, video gastroscope

Why the product was recalled

If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.

Distribution information

Distributed US (nationwide) and in Australia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.