FDA recall record

Olympus ENDOEYE HD II Video Telescopes Models: WA50040A Endoeye HD II, 10 mm, 0 deg. recall — FDA Z-1929-2017

A damaged temperature sensor at the endoscope's tip can cause the endoscope's distal end to become abnormally hot.

Class IIclassification
Z-1929-2017official record ID
May 10, 2017report date

Recall record at a glance

Official recordZ-1929-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOlympus Corporation of the Americas
Report dateMay 10, 2017
Recall initiation dateFeb 10, 2017
Quantity863 units
LocationCenter Valley, PA, United States

Official product description

Olympus ENDOEYE HD II Video Telescopes Models: WA50040A Endoeye HD II, 10 mm, 0 deg., autoclavable; WA50042A Endoeye HD II, 10 mm, 30 deg., autoclavable; repair cable 10mm 0/30 deg., part number W7100603

Why the product was recalled

A damaged temperature sensor at the endoscope's tip can cause the endoscope's distal end to become abnormally hot.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.