FDA recall record

GE OEC Medical Systems OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile recall — FDA Z-0706-2025

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Class IIclassification
Z-0706-2025official record ID
Dec 25, 2024report date

Recall record at a glance

Official recordZ-0706-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE OEC Medical Systems, Inc.
Report dateDec 25, 2024
Recall initiation dateNov 21, 2024
Quantity25 systems
LocationSalt Lake City, UT, United States

Official product description

OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

Why the product was recalled

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Distribution information

U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.