FDA recall record

Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen recall — FDA Z-1805-2020

During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can…

Class IIclassification
Z-1805-2020official record ID
May 6, 2020report date

Recall record at a glance

Official recordZ-1805-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmObalon Therapeutics Inc.
Report dateMay 6, 2020
Recall initiation dateMay 23, 2019
Quantity45 units
LocationCarlsbad, CA, United States

Official product description

Obalon Balloon System with Model 4300 Touch Dispenser Touchscreen; Software version 02.00.00.0076 - Product Usage: The System is intended to be used as an adjunct to a moderate intensity diet and behavior modification program. All balloons must be removed 6 months after the first balloon is placed. The Obalon Balloons are administered using the Navigation System.

Why the product was recalled

During initial set-up and/or replacement of the dispenser batteries, the touch dispenser touchscreen can inadvertently lose calibration. This issue can also occur if the touchscreen is pressed when the device is powered on in preparation for a balloon administration. As a result, the touchscreen can become unresponsive and balloon inflation cannot be initiated.

Distribution information

US Nationwide distribution including the states of VA, NY, GA, TX, TN, OR, FL, LA, CT, and NJ.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.