FDA recall record

NxStage System One S Cycler -High Permeability Hemodialysis System Model no recall — FDA Z-0337-2016

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF…

Class IIclassification
Z-0337-2016official record ID
Dec 2, 2015report date

Recall record at a glance

Official recordZ-0337-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNxStage Medical, Inc.
Report dateDec 2, 2015
Recall initiation dateOct 29, 2015
Quantity7 units
LocationLawrence, MA, United States

Official product description

NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4

Why the product was recalled

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Distribution information

Worldwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 4, 2025

NxStage PureFlow-B Solution recall — FDA Z-1864-2025

Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam…

Class IIFDA: OngoingNxStage Medical Inc.