FDA recall record

NxStage Express Fluid Warmer recall — FDA Z-2793-2018

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is…

Class IIclassification
Z-2793-2018official record ID
Aug 22, 2018report date

Recall record at a glance

Official recordZ-2793-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNxStage Medical, Inc.
Report dateAug 22, 2018
Recall initiation dateJun 11, 2018
Quantity14975
LocationLawrence, MA, United States

Official product description

NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessory to the NxStage System One used to warm therapy fluids prior to administration.

Why the product was recalled

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

Distribution information

Worldwide Distribution - US Nationwide in the countries of Australia, Canada, France, Germany, Italy, Kuwait, Netherlands, Saudi Arabia, Spain, Sweden, and UK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 4, 2025

NxStage PureFlow-B Solution recall — FDA Z-1864-2025

Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam…

Class IIFDA: OngoingNxStage Medical Inc.