Recall record at a glance
| Official record | Z-0831-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Completed |
| Recalling firm | Nuwellis Inc. |
| Report date | Jan 22, 2025 |
| Recall initiation date | Dec 11, 2024 |
| Quantity | 845 units |
| Location | Eden Prairie, MN, United States |
Official product description
nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System
Why the product was recalled
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
Distribution information
US Nationwide Distribution and the country of Hong Kong
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.