Recall record at a glance
| Official record | Z-1350-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Nuvasive Specialized Orthopedics Inc. |
| Report date | Apr 14, 2021 |
| Recall initiation date | Feb 12, 2021 |
| Quantity | 229 units |
| Location | Aliso Viejo, CA, United States |
Official product description
NUVASIVE PRECICE UNYTE System (Humerus and Tibia): PRECICE Trauma Nail System/PRECICE Unyte-Humerus Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 165 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 180 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 195 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 210 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 225 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 240 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 255 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 270 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 285 mm length; Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 300 mm length; End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 0mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 2mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 5mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 10mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 15mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 0mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 2mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 5mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 10mm); End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 15mm); 3.5 mm Locking Screw, 20 mm Length; 3.5 mm Locking Screw 25 mm Length; 3.5 mm Locking Screw 30 mm Length; 3.5 mm Locking Screw 35 mm Length; 3.5 mm Locking Screw 40 mm Length; 3.5 mm Locking Screw 45 mm Length; 3.5 mm Locking Screw 50 mm Length; 3.5 mm Locking Screw 55 mm Length; 3.5 mm Locking Screw 60 mm Length; Locking Screw, 3.5x65mm; Locking Screw, 3.5x70mm; Locking Screw, 3.5x75mm; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm Locking Screw 20 mm Length; 5.0 mm Locking Screw 25 mm Length; 5.0 mm Locking Screw 30 mm Length; 5.0 mm Locking Screw 35 mm Length; 5.0 mm Locking Screw 40 mm Length; 5.0 mm Locking Screw 45 mm Length; 5.0 mm Locking Screw 50 mm Length; 5.0 mm Locking Screw 55 mm Length; 5.0 mm Locking Screw 60 mm Length; 5.0 mm Locking Screw 65 mm Length; 5.0 mm Locking Screw 70 mm Length; 5.0 mm Locking Screw 75 mm Length; 5.0 mm X 80 mm Locking Screw, Peg; 5.0 mm X 85 mm Locking Screw, Peg; 5.0 mm X 90 mm Locking Screw, Peg; 5.0 mm X 95 mm Locking Screw, Peg; 5.0 mm X 100 mm Locking Screw, Peg; PRECICE Trauma Nail System/Precice Unyte-Tibia Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 240 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 255 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 270 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 285 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 300 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 315 mm length; Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 330 mm length; Orthopedic trauma intra
Why the product was recalled
Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.
Distribution information
Worldwide distribution, including U.S. Nationwide.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.