Recall record at a glance
| Official record | Z-2684-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | NuVasive Inc. |
| Report date | Jul 29, 2020 |
| Recall initiation date | Jun 10, 2020 |
| Quantity | 23 |
| Location | San Diego, CA, United States |
Official product description
Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal surgery.
Why the product was recalled
The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5.
Distribution information
US Nationwide distribution including in the states of AR, CA, IA, IL, WI. OUS: None
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.