Recall record at a glance
| Official record | Z-1097-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Collagen Matrix Inc. |
| Report date | Mar 23, 2016 |
| Recall initiation date | May 13, 2013 |
| Quantity | 347 units |
| Location | Franklin Lakes, NJ, United States |
Official product description
NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) -Filling of pre-implant defects in conjunction with products intended for Guided Bone Regeneration
Why the product was recalled
During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.
Distribution information
US Distribution to state of: MA.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.