Recall record at a glance
| Official record | Z-2869-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | In2Bones, SAS |
| Report date | Aug 9, 2017 |
| Recall initiation date | Jul 13, 2017 |
| Quantity | 287 screws |
| Location | Ecully, N/A, France |
Official product description
Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy
Why the product was recalled
The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis
Distribution information
Worldwide Diribution - US including Tennessee; Foreign - Italy, Portugal, Spain, Switzerland, UK, and France
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: N/A.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.