FDA recall record

Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII recall — FDA Z-0223-2015

Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event…

Class IIclassification
Z-0223-2015official record ID
Nov 26, 2014report date

Recall record at a glance

Official recordZ-0223-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNidek Inc.
Report dateNov 26, 2014
Recall initiation dateAug 6, 2012
Quantity14 units
LocationFremont, CA, United States

Official product description

Nidek Excimer Laser Corneal Surgery System, model EC-5000 CXIII. Opthalmic Device. The system is composed as a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system and a computer for system control.

Why the product was recalled

Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.

Distribution information

US Distribution including the states of CO, NY, MO, LA, ME, CA, MI, AZ, CT and FL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.