FDA recall record

Natus Neurology Nicolet EEG Wireless Amplifier 32/64 Channel recall — FDA Z-0946-2013

Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the…

Class IIclassification
Z-0946-2013official record ID
Mar 20, 2013report date

Recall record at a glance

Official recordZ-0946-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNatus Neurology Inc.
Report dateMar 20, 2013
Recall initiation dateFeb 27, 2013
Quantity114 (31 USA; 83 OUS)
LocationMiddleton, WI, United States

Official product description

Nicolet EEG Wireless Amplifier 32/64 Channel. Catalog/Part Numbers : 515-015400 for the 32 channel amplifier and 515-015500 for the 64 channel amplifier. Product Usage: The Nicolet Wireless EEG Amplifier is intended to be used as a front end amplifier to acquire, store, and transmit electrophysiological signals in a wired or wireless mode for the Nicolet Neurodiagnostic system.

Why the product was recalled

Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the on/off button. The overheating may result in discomfort ranging from too hot up to the potential for a burn.

Distribution information

Worldwide Distribution - USA Nationwide including the states of CA, FL, LA, MD, MS, MO, NJ, NY, PA, TX and the country of AUSTRALIA. BAHRAIN BRAZIL CANADA Country DENMARK EGYPT GERMANY HONG KONG INDIA INDONESIA IRELAND ITALY JAPAN KOREA KUWAIT MALAYSIA MEXICO NETHERLANDS ROMANIA RUSSIA SINGAPORE SPAIN SWITZERLAND TAIWAN THAILAND TURKEY UNITED KINGDOM URUGUAY VIET NAM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.