FDA recall record

Nextremity Solutions Nextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS recall — FDA Z-1458-2020

Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from…

Class IIclassification
Z-1458-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1458-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNextremity Solutions
Report dateMar 18, 2020
Recall initiation dateFeb 12, 2020
Quantity190 units
LocationWarsaw, IN, United States

Official product description

Nextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS, Sterile. REF: NX-4532K - Product Usage: The Nextra Hammertoe Correction System is a system of implants and associated instruments which is indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes.

Why the product was recalled

Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant

Distribution information

International distribution in the countries of Spain, United Kingdom, Germany, and Switzerland.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.