FDA recall record

Zimmer NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer recall — FDA Z-1433-2025

The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the…

Class IIclassification
Z-1433-2025official record ID
Apr 2, 2025report date

Recall record at a glance

Official recordZ-1433-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmZimmer, Inc.
Report dateApr 2, 2025
Recall initiation dateFeb 27, 2025
Quantity76 units
LocationWarsaw, IN, United States

Official product description

NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10

Why the product was recalled

The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.

Distribution information

Worldwide - US Nationwide distribution in the states pf CA, KS, MA, MD, ME, NE, NJ, OR, SD, and TX. The countries of Canada, China, India, Netherlands, Singapore.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0972-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0970-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0975-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0968-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.