FDA recall record

Zimmer NexGen Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial recall — FDA Z-0480-2014

Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the…

Class IIclassification
Z-0480-2014official record ID
Dec 18, 2013report date

Recall record at a glance

Official recordZ-0480-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZimmer, Inc.
Report dateDec 18, 2013
Recall initiation dateOct 31, 2013
Quantity128 units
LocationWarsaw, IN, United States

Official product description

NexGen Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis. The device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis; Collagen disorders, and/or avascular necrosis of the femoral condyle; Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, varus, or flexion deformities.

Why the product was recalled

Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial plate would not hold the stem as the connection was too loose. The taper diameter was found to be oversized by .010.

Distribution information

Worldwide Distribution - USA (nationwide) including the states of CA, IN, OH, GA, WI, IA, NE, NJ, PA, NY, MD, NC, AZ, NM, TX, CO, IL, AL, MI, WA, FL and KS, and the countries of China, Japan and Taiwan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0972-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0970-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0975-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0968-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.