FDA recall record

Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 recall — FDA Z-1874-2017

Medtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown…

Class Iclassification
Z-1874-2017official record ID
May 10, 2017report date

Recall record at a glance

Official recordZ-1874-2017
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmNewport Medical Instruments Inc.
Report dateMay 10, 2017
Recall initiation dateMar 30, 2017
Quantity12,966
LocationCosta Mesa, CA, United States

Official product description

Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs). The Newport HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The Newport HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Why the product was recalled

Medtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.

Distribution information

Worldwide Distribution - US (Nationwide) Internationally to Canada, Australia, Brazil, Japan, Korea, Mexico, Bahrain, Egypt, Greece Hungary, Ireland, Israel, Jordan, Kazakhstan, Lithuania, Mauritius Namibia, Norway, Pakistan, Poland, Qatar, Romania, Russian Federation South Africa, Spain, Turkey, United Arab Emirates, United Kingdom Algeria, Armenia, Austria, Bahrain, Belarus, Belgium, Botswana Bulgaria, Cyprus, Czech Republic, Denmark, Egypt, Finland, France Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Kenya, Lebanon Lithuania, Macedonia, Malawi, Malta, Mauritius, Mozambique, Netherlands Norway, Oman, Poland, Portugal, Qatar, Republic of Moldova, Romania Saudi Arabia, Serbia, Slovakia, Sudan, Sweden, Switzerland, Tunisia Turkey, Uganda, Ukraine, United Kingdom, Yemen, and Zimbabwe

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.