FDA recall record

Neusoft Medical Systems Co NeuViz 128 Multi-slice CT Scanner System with software version recall — FDA Z-0876-2017

The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be…

Class IIclassification
Z-0876-2017official record ID
Jan 4, 2017report date

Recall record at a glance

Official recordZ-0876-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNeusoft Medical Systems Co., Ltd.
Report dateJan 4, 2017
Recall initiation dateAug 31, 2016
Quantity23 units
LocationShenyang, N/A, China

Official product description

NeuViz 128 Multi-slice CT Scanner System with software version 1.0.7.4038+P06 or previous version

Why the product was recalled

The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.

Distribution information

US Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.