| Official record | Z-2093-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Ongoing |
| Recalling firm | NeuMoDx Molecular Inc. |
| Report date | Jul 12, 2023 |
| Recall initiation date | May 15, 2023 |
| Quantity | 1,086 test strips |
| Location | Ann Arbor, MI, United States |
Official product description
NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs in transport medium and bronchoalveolar lavage (BAL) specimens from individuals suspected of COVID-19 by their healthcare provider (HCP). REF Number: 300800
Why the product was recalled
Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.
Distribution information
US Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.