FDA recall record

Instradent USA NEODENT CM Intraoral Scanbody, Article Number: 108.139 recall — FDA Z-1831-2017

During internal testing, Neodent determined that the body diameter (3.8mm) of the Neodent 108.139 CM Exact Intraoral Scan body is not compatible with the…

Class IIclassification
Z-1831-2017official record ID
Apr 26, 2017report date

Recall record at a glance

Official recordZ-1831-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInstradent USA, Inc.
Report dateApr 26, 2017
Recall initiation dateMar 13, 2017
Quantity170 pieces
LocationAndover, MA, United States

Official product description

NEODENT CM Intraoral Scanbody, Article Number: 108.139

Why the product was recalled

During internal testing, Neodent determined that the body diameter (3.8mm) of the Neodent 108.139 CM Exact Intraoral Scan body is not compatible with the platform diameter of CM Implants when the implant is intraosseoussly (subcrestally) positioned.

Distribution information

Distributed to the states of: NJ, TX, NV, CA, PA, AZ, and IL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.