FDA recall record

Mercury Enterprises, Inc. dba Mercury Medical Neo-Tee T-Piece Resuscitator, for pediatric recall — FDA Z-0821-2025

Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high…

Class Iclassification
Z-0821-2025official record ID
Jan 22, 2025report date

Recall record at a glance

Official recordZ-0821-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMercury Enterprises, Inc. dba Mercury Medical
Report dateJan 22, 2025
Recall initiation dateDec 10, 2024
Quantity37370
LocationClearwater, FL, United States

Official product description

Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842

Why the product was recalled

Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.

Distribution information

Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.