FDA recall record

Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 recall — FDA Z-1412-2015

neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket…

Class IIclassification
Z-1412-2015official record ID
Apr 22, 2015report date

Recall record at a glance

Official recordZ-1412-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNatus Medical Inc.
Report dateApr 22, 2015
Recall initiation dateMar 16, 2015
Quantity2611 units (1911 units in the US and 700 units outside the US)
LocationSeattle, WA, United States

Official product description

Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS

Why the product was recalled

neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Distribution information

distributed worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records