FDA recall record

AngioDynamics NAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. recall — FDA Z-1346-2016

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of…

Class IIclassification
Z-1346-2016official record ID
Apr 13, 2016report date

Recall record at a glance

Official recordZ-1346-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAngioDynamics Inc.
Report dateApr 13, 2016
Recall initiation dateNov 3, 2015
Quantity45 kits
LocationGlens Falls, NY, United States

Official product description

NAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. 601319810, UPN Product No. H7496013198101 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire.

Why the product was recalled

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Distribution information

US Nationwide in the states of CA, IL, MA, SC, TX

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.