FDA recall record

Alliance Partners Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill recall — FDA Z-1989-2015

Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw…

Class IIclassification
Z-1989-2015official record ID
Jul 15, 2015report date

Recall record at a glance

Official recordZ-1989-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAlliance Partners LLC
Report dateJul 15, 2015
Recall initiation dateJun 9, 2015
Quantity21 units
LocationSan Antonio, TX, United States

Official product description

Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide Product Usage: The Nakoma-SL ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patients with the following indications: Degenerative Disk Disease, Trauma, Tumors, Deformities or curvatures, Psuedoarthrosis, Failed previous fusions, Spondylolistesis, and Spinal Stenosis.

Why the product was recalled

Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.

Distribution information

US Nationwide Distribution in the states LA, TX, and PR.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.