Recall record at a glance
| Official record | Z-0708-2014 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | The Anspach Effort, Inc. |
| Report date | Jan 15, 2014 |
| Recall initiation date | Oct 26, 2012 |
| Quantity | 6,824 imits (4,7324 in USA / 2,100 Foreign) |
| Location | Palm Beach Gardens, FL, United States |
Official product description
Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD 1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM. The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting.
Why the product was recalled
During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu
Distribution information
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.