Recall record at a glance
| Official record | Z-0647-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Terminated |
| Recalling firm | Centurion Medical Products Corporation |
| Report date | Dec 14, 2016 |
| Recall initiation date | Oct 21, 2016 |
| Quantity | 1000 kits |
| Location | Williamston, MI, United States |
Official product description
Multi-Med Single Lumen Catheters sold in Centurion convenience kits. The Multi-Med catheter is indicated for use in patients requiring administration of solutions, blood sampling, and central venous pressure monitoring.
Why the product was recalled
The kits contain Multi-Med Single Lumen Catheters that have a potential for excess material to remain at the tip of the catheter from the manufacturing process. It is possible that the excess material may separate from the catheter during use and pose a potential risk of embolism, which could result in serious injury or death.
Distribution information
Nationwide Distribution
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.