FDA recall record

Siemens Medical Solutions USA Multi Display Manager in the Artis Systems recall — FDA Z-2302-2020

Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the…

Class IIclassification
Z-2302-2020official record ID
Jun 17, 2020report date

Recall record at a glance

Official recordZ-2302-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateJun 17, 2020
Recall initiation dateApr 30, 2020
Quantity15 US
LocationMalvern, PA, United States

Official product description

Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280; Artis Q ceiling Model: 10848281; Artis Q biplane Model:10848282; Artis zee ceiling Model: 10094137; Artis zee floor Model: 10094135 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Why the product was recalled

Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.