FDA recall record

Philips MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152 recall — FDA Z-2213-2025

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or…

Class IIclassification
Z-2213-2025official record ID
Aug 6, 2025report date

Recall record at a glance

Official recordZ-2213-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPhilips North America
Report dateAug 6, 2025
Recall initiation dateJun 30, 2025
Quantity551 units
LocationCambridge, MA, United States

Official product description

MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;

Why the product was recalled

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Distribution information

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, French Guiana, FrenchPolynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen, Zambia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.