FDA recall record

Orthoscan Mobile Mini C-arm system Part # 1000-0004 recall — FDA Z-0259-2018

OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a…

Class IIclassification
Z-0259-2018official record ID
Jan 17, 2018report date

Recall record at a glance

Official recordZ-0259-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOrthoscan, Inc.
Report dateJan 17, 2018
Recall initiation dateAug 31, 2017
Quantity2012 total
LocationScottsdale, AZ, United States

Official product description

Mobile Mini C-arm system Part # 1000-0004. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.

Why the product was recalled

OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user

Distribution information

Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 11, 2024

Orthoscan Ziehm Vision RFD 3D recall — FDA Z-0579-2025

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR…

Class IIFDA: OngoingOrthoscan, Inc.