FDA recall record

Abbott Vascular MitraClip Clip Delivery System, product number MSK0101 recall — FDA Z-0911-2016

Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted…

Class Iclassification
Z-0911-2016official record ID
Mar 16, 2016report date

Recall record at a glance

Official recordZ-0911-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmAbbott Vascular
Report dateMar 16, 2016
Recall initiation dateFeb 4, 2016
Quantity7,600 units total (2,300 units in US)
LocationTemecula, CA, United States

Official product description

MitraClip Clip Delivery System, product number MSK0101. The MitraClip System contains the Clip Delivery System, product number CDS0201, GTIN 08717648195914 and the Steerable Guide Catheter, product number SGC0101, GTIN 08717648195921. The MitraClip Clip Delivery System (CDS) consists of three major components: 1) the Delivery Catheter 2) the Steerable Sleeve, and 3) the MitraClip device. The implantable MitraClip device is located at the distal end of the CDS. The CDS is used to advance and manipulate the implantable MitraClip device for proper positioning and placement on the mitral valve leaflets. The Delivery Catheter, which is part of the CDS, is designed to deliver and deploy the MitraClip. The Delivery Catheter extends through the thru-lumen of the Steerable Sleeve.

Why the product was recalled

Abbott Vascular has recently received reports of cases on Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitraClip, but the Clip could not be detached from the delivery system due to a mandrel fracture.

Distribution information

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Argentina Luxembourg Australia Malaysia Austria Mexico Belgium Netherlands Brazil New Zealand Canada Norway Colombia Philippines Corrientes Poland Czech Republic Portugal Denmark Egypt Romania Finland Saudi Arabia France Slovenia Germany Spain Greece Sweden Hong Kong Switzerland Hungary Thailand Indonesia Turkey Israel United Arab Emirates Italy United Kingdom Kuwait and Vietnam.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.