FDA recall record

Medtronic MiniMed MiniMed Insulin Pump Device recall — FDA Z-0003-2025

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced…

Class Iclassification
Z-0003-2025official record ID
Oct 9, 2024report date

Recall record at a glance

Official recordZ-0003-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMedtronic MiniMed, Inc.
Report dateOct 9, 2024
Recall initiation dateJul 31, 2024
Quantity526,558
LocationNorthridge, CA, United States

Official product description

MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896

Why the product was recalled

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Distribution information

US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.