Recall record at a glance
| Official record | Z-1635-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Medtronic MiniMed, Inc. |
| Report date | Apr 1, 2026 |
| Recall initiation date | Jan 30, 2026 |
| Quantity | 26,851 pumps |
| Location | Northridge, CA, United States |
Official product description
MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
Why the product was recalled
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
Distribution information
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH,NJ, NM, NV, NY, OH,OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Israel.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.