FDA recall record

Medtronic Minimed MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ recall — FDA Z-1772-2021

Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the…

Class IIclassification
Z-1772-2021official record ID
Jun 9, 2021report date

Recall record at a glance

Official recordZ-1772-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedtronic Minimed
Report dateJun 9, 2021
Recall initiation dateMar 15, 2021
QuantityU.S. Investigational Clinical Study = 1,266 devices
LocationNorthridge, CA, United States

Official product description

MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 670G V6.3 SF MG CLIN FR;¿ PUMP MMT-1741KFX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KX 670G V6.3 SF MG CLIN GER;¿ PUMP MMT-1741KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KXS 670G V6.3 SF MG CLIN IS;¿ PUMP MMT-1741KXS 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1742KX 670G V6.1 MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN SLO;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 MM;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 EU MM;¿

Why the product was recalled

Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.

Distribution information

U.S.(Clinical): AR, CA, CO, CT, FL, GA, ID, MI, MN, TN, TX, and WA O.U.S.: Austria, Bahrain, Belgium, Canary Islands, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Liechtenstein, Luxembourg, Malta, Martinique, Netherlands, New Caledonia, Norway, Poland, Qatar, Reunion, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.