FDA recall record

Medtronic MiniMed MiniMed 620G, 3.0mL mg/dL, Model No recall — FDA Z-0437-2016

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being…

Class IIclassification
Z-0437-2016official record ID
Dec 23, 2015report date

Recall record at a glance

Official recordZ-0437-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic MiniMed Inc.
Report dateDec 23, 2015
Recall initiation dateSep 25, 2015
Quantity53 units
LocationNorthridge, CA, United States

Official product description

MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Why the product was recalled

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

Distribution information

Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.