FDA recall record

Angiodynamics MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the recall — FDA Z-1343-2024

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass…

Class IIclassification
Z-1343-2024official record ID
Mar 27, 2024report date

Recall record at a glance

Official recordZ-1343-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAngiodynamics, Inc.
Report dateMar 27, 2024
Recall initiation dateJan 30, 2024
Quantity130 units
LocationQueensbury, NY, United States

Official product description

MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-765

Why the product was recalled

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Distribution information

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.