FDA recall record

Beckman Coulter MicroScan WalkAway-96 plus Reconditioned Instrument REF B1018-396R recall — FDA Z-1088-2022

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the…

Class IIclassification
Z-1088-2022official record ID
May 25, 2022report date

Recall record at a glance

Official recordZ-1088-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Inc.
Report dateMay 25, 2022
Recall initiation dateOct 21, 2019
Quantity4 systems
LocationBrea, CA, United States

Official product description

MicroScan WalkAway-96 plus Reconditioned Instrument REF B1018-396R

Why the product was recalled

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

Distribution information

Worldwide distribution - U.S. Nationwide distribution in the states of Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Pennsylvania, Tennessee, Texas, Utah, Vermont, Virginia, and Washington. The countries of Belgium, Brazil, Colombia, Germany, Hong Kong, Indonesia, Italy, Japan, Paraguay, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, United Arab Emirates, Vietnam and Yemen.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.