FDA recall record

MEVION S250/MEVION S250i Proton Beam Radiation therapy system Product Usage: Indicated for recall — FDA Z-2746-2018

QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The…

Class IIclassification
Z-2746-2018official record ID
Aug 22, 2018report date

Recall record at a glance

Official recordZ-2746-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMevion Medical Systems, Inc.
Report dateAug 22, 2018
Recall initiation dateJun 29, 2018
Quantity7
LocationLittleton, MA, United States

Official product description

MEVION S250/MEVION S250i Proton Beam Radiation therapy system Product Usage: Indicated for the delivery of radiation for the treatment of patients with localized tumors or other conditions susceptible to treatment by radiation.

Why the product was recalled

QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that can be damaged when used with incompatible (third-party) accessories. This can create unsafe conditions for use and may compromise the subsequent use of any kVue inserts, whether authorized or not. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury.

Distribution information

US Nationwide Distribution in the states of DC, FL, MO, NJ, OH, and OK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 2, 2020

MEVION S250i, MEVION S250 recall — FDA Z-0485-2021

Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur

Class IIFDA: TerminatedMevion Medical Systems, Inc.
Medical DeviceJul 1, 2020

MEVION S250i, MEVION S250 recall — FDA Z-2406-2020

Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off

Class IIFDA: TerminatedMevion Medical Systems, Inc.