FDA recall record

St Jude Medical Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models recall — FDA Z-0033-2018

New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF)…

Class IIclassification
Z-0033-2018official record ID
Oct 25, 2017report date

Recall record at a glance

Official recordZ-0033-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSt Jude Medical Inc.
Report dateOct 25, 2017
Recall initiation dateAug 28, 2017
Quantity495,011 units (391,093 in US)
LocationSylmar, CA, United States

Official product description

Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.

Why the product was recalled

New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.

Distribution information

Worldwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.