FDA recall record

Medtronic SynchroMed II Implantable Drug Infusion Pump, Model 8637-20, 8637-40 recall — FDA Z-1570-2014

This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can…

Class IIclassification
Z-1570-2014official record ID
May 14, 2014report date

Recall record at a glance

Official recordZ-1570-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Neuromodulation
Report dateMay 14, 2014
Recall initiation dateFeb 26, 2014
Quantity195,198 pumps (146,435 US, 48,763 OUS)
LocationMinneapolis, MN, United States

Official product description

Medtronic SynchroMed II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.

Why the product was recalled

This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re

Distribution information

Worldwide Distribution - All states in USA. OUS: List not provided at this time.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.