FDA recall record

Medtronic Strata II / Strata NSC Valves recall — FDA Z-1121-2017

Medtronic Neurosurgery initiated a recall after receiving five (5) Product Experiences between February 15, 2015 and August 23, 2016 reporting of…

Class IIclassification
Z-1121-2017official record ID
Feb 8, 2017report date

Recall record at a glance

Official recordZ-1121-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Neurosurgery
Report dateFeb 8, 2017
Recall initiation dateJan 17, 2017
Quantity315,997
LocationGoleta, CA, United States

Official product description

Medtronic Strata II / Strata NSC Valves (includes Burr Hole and Lumbar Peritoneal. All Catalog Numbers. Strata-type valves are used in the management of hydrocephalus. The Strata II/Strata NSC valve is a shunt component designed to allow for continuous flow of cerebrospinal fluid (CSF) from the ventricles of the brain to the peritoneal cavity or right atrium of the heart. mechanism. Strata Valve Magnet Reverse Polarity (Excludes StrataMR). Catalog No. 27739 27740 27786 27787 27788 27789 27794 27812 27814 27815 27816 27817 27818 27820 27821 27822 27823 27824 27825 27827 27828 27830 27831 27832 27848 27864 27867 27868 27888 27903 27922 27823 27924 23042 27925 27926 27927 27932 27933 42335 42355 42365 42836 42856 42866 44420 44421 44430 44465 46070 46075 46080 46085 46635 46655 46665 46836 46856 46866 46871 46876 46881 46886 92355 92365 92856 92866 96655 96665

Why the product was recalled

Medtronic Neurosurgery initiated a recall after receiving five (5) Product Experiences between February 15, 2015 and August 23, 2016 reporting of discrepancies between the pressure/performance level (PL) at which the valve was set and the readings determined via an x-ray image.

Distribution information

Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Algeria India Poland Armenia Indonesia Portugal Australia Iran Austria Iraq Qatar Bangladesh Ireland Republic of Montenegro Barbados Israel Republic of Serbia Belarus Italy Romania Belgium Jamaica Russian Federation Bosnia and Herzegovina Japan Saudi Arabia Botswana Jordan Singapore Bulgaria Kazakhstan Slovakia Cambodia Kenya South Africa Canada Kuwait South Korea Canary Islands Latvia Spain Cayman Islands Lebanon Sri Lanka China Libya Sweden Croatia Lithuania Switzerland Cyprus Macau Syria

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records