FDA recall record

Medtronic recall — FDA Z-1617-2021

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor…

Class IIclassification
Z-1617-2021official record ID
May 26, 2021report date

Recall record at a glance

Official recordZ-1617-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedtronic Inc.
Report dateMay 26, 2021
Recall initiation dateFeb 26, 2021
Quantity1,174 units
LocationNorthridge, CA, United States

Official product description

Medtronic, REF: MMT-7008C, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 5PK/ GTIN: 20763000252032 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Why the product was recalled

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Distribution information

Worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.