FDA recall record

Micro Therapeutics Medtronic Pipeline Vantage Embolization Device with Shield Technology recall — FDA Z-1246-2025

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation…

Class Iclassification
Z-1246-2025official record ID
Mar 12, 2025report date

Recall record at a glance

Official recordZ-1246-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMicro Therapeutics, Inc.
Report dateMar 12, 2025
Recall initiation dateJan 30, 2025
Quantity7820
LocationIrvine, CA, United States

Official product description

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2

Why the product was recalled

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

Distribution information

Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.