FDA recall record

Heartware Medtronic HVAD Pump Implant Kits recall — FDA Z-2478-2021

There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than…

Class IIclassification
Z-2478-2021official record ID
Sep 22, 2021report date

Recall record at a glance

Official recordZ-2478-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHeartware, Inc.
Report dateSep 22, 2021
Recall initiation dateAug 6, 2021
Quantity11851 (potentially active, US); 1995 (active, US)
LocationMiami Lakes, FL, United States

Official product description

Medtronic HVAD Pump Implant Kits

Why the product was recalled

There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.

Distribution information

Distribution throughout the US. Also distribution in ARGENTINA Armenia, Australia, Austria, BELGIUM, BRAZIL, CANADA, CHILE, Colombia, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, Germany, GREECE HONG KONG, HUNGARY, INDIA, ISRAEL, Italy, JAPAN, Kazakhstan, Korea, KUWAIT, Latvia, LEBANON, LITHUANIA, LUXEMBOURG, Macedonia, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, Poland, ROMANIA, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab Emirates, United Kingdom, VIETNAM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 13, 2023

Heartware PUMP 1103 HVAD recall — FDA Z-0478-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0482-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0481-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0475-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0479-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP MCS1705PU HVAD recall — FDA Z-0483-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.