FDA recall record

Medtronic Neuromodulation Medtronic, External neurostimulator, ENS (37021) & ENS (37022) recall — FDA Z-1526-2014

Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be…

Class IIclassification
Z-1526-2014official record ID
May 7, 2014report date

Recall record at a glance

Official recordZ-1526-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Neuromodulation
Report dateMay 7, 2014
Recall initiation dateFeb 27, 2014
Quantity170,286 US, 73,921 OUS for all devices.
LocationMinneapolis, MN, United States

Official product description

Medtronic, External neurostimulator, ENS (37021) & ENS (37022). Including the external neurostimulators which are associated with software application run on the NVision Clinician Programmer and used during trialing for spinal cord stimulation (SCS) patients.

Why the product was recalled

Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval

Distribution information

Nationwide Distribution - all states including DC and Puerto Rico.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.