FDA recall record

Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in recall — FDA Z-3212-2018

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for…

Class IIclassification
Z-3212-2018official record ID
Oct 3, 2018report date

Recall record at a glance

Official recordZ-3212-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Neurosurgery
Report dateOct 3, 2018
Recall initiation dateMay 3, 2018
Quantity2616
LocationGoleta, CA, United States

Official product description

Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use

Why the product was recalled

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.

Distribution information

Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records