Recall record at a glance
| Official record | Z-2603-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Medtronic Sofamor Danek USA Inc. |
| Report date | Aug 15, 2018 |
| Recall initiation date | Jul 9, 2018 |
| Quantity | 47 units |
| Location | Memphis, TN, United States |
Official product description
Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.MULTI-AXIAL SCREW FOR 5.5/6.0MM RODS, REF 55840005045
Why the product was recalled
This lot has incorrectly been laser marked both on the screw head itself as well as the data carrier tag attached to the screw. The description on face two of the screw head and on the data carrier tag both incorrectly indicate a 5.5 mm diameter screw when the actual screw diameter aligns with 5.0 diameter screw size requirements.
Distribution information
US
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.